Home | Jobs | News & Articles | Job Advice | Search | Protocols | Fun | @RealLabRat
Lab Rat with DNA

Post your resume
Candidates - post your resume

Search job listings
Candidates - search biotech jobs

Post job openings
Employers - post job openings

Email Login
Password
New users
sign up!
HOME > Protocols > Cell Biology > Protocol for ENLITEN® ATP Assay System Bioluminescence Detection Kit for ATP

Protocol for ENLITEN® ATP Assay System Bioluminescence Detection Kit for ATP

1. INTRODUCTION

The Turner BioSystems TD-20/20 Luminometer combined with the Promega ENLITEN® ATP Assay System Bioluminescence Detection Kit provides a sensitive, rapid method for measuring adenosine 5'-triphosphate (ATP). A reliable method for ATP detection is useful for studying enzymes that produce or degrade ATP. ATP detection also provides an indirect measurement of microbes, food residue, or other biological material.

The ATP-dependent oxidation of luciferin by luciferase produces light. When ATP is the limiting factor in the luciferin oxidation reaction, the amount of light produced is proportional to the ATP concentration of the sample.

The highly sensitive TD-20/20 can detect attomole levels of ATP. The limit of detection for ATP with the ENLITEN® detection assay is 1x10-16 moles (Figure 1).

Figure 1. A standard curve obtained using the TD-20/20 Luminometer and the ENLITEN® ATP Assay System.
10 µL of ATP Standard diluted in HEPES buffer was added to a test tube containing 100 µL rL/L Reagent.

When designing your ATP assay with the ENLITEN® kit, it is important to remember several key aspects of the luciferase reaction. First, the rL/L reacts optimally at a pH of 7.73 and 23—25°C. Salts and many nonionic chemicals will impair light production. Therefore, it is necessary to take care when selecting buffers and ATP extractants for the sample preparations. It is also recommended to check for ATP contamination in your assay buffer by comparing the RLU values obtained with your assay buffer and rL/L to those of ATP-Free Water.

Within your sample, there may exist several different ATP stores. For example, in cell preparations, ATP may be present in the media. Treatment of cells may alter the amount of ATP in the media.

If you wish to measure the ATP content in microorganisms or cells, you will need to extract the ATP before analysis. Trichloroacetic acid (TCA) is recommended because it releases ATP from cells and inactivates ATP-degrading enzymes. Because TCA inhibits the luciferase reaction, it is important to determine the minimum amount of TCA necessary. Generally, 0.5% to 2.5% TCA (final concentration) is sufficient for ATP extraction from bacteria and eukaryotic cells.

Preparing a standard curve is a useful tool for proper ATP analysis of your samples. A standard curve should be prepared daily or whenever a new aliquot of the rL/L Reagent is used. The standard curve should include varying concentrations of ATP diluted in your assay buffer. The composition of this buffer should be identical to the composition of the buffer used in your assay to prepare your samples.

Keeping these considerations in mind will help you obtain the most accurate ATP analysis.

2. MATERIALS REQUIRED

From Turner BioSystems:

  • TD-20/20 Luminometer (P/N 2020-000)
  • 12 mm Test Tube Holder (P/N 2020-950)
  • 12 x 50 mm Polypropylene Test Tubes (P/N 2020-955)

From Promega:

  • ENLITEN® ATP Assay System Bioluminescence Detection Kit for ATP (P/N FF2000)
    • Contains:
      • 1 vial rLuciferase/Luciferin (rL/L)Reagent
      • 12 mL Reconstitution Buffer
      • 1 vial ATP Standard (1x10-7 M)
      • 25 mL ATP-free Water

Other Materials:

  • Adjustable p1000 Volume Pipetter and Tips
  • Adjustable p200 Volume Pipetter and Tips
  • Assay buffer
  • Test tube rack compatible with 12 x 50 mm polypropylene tubes
  • Nitrile, vinyl, or latex gloves

Storage Conditions: The rL/L Reagent and Reconstitution Buffer must be stored at -20°C prior to reconstitution. Store the ATP Standard at -20°C.

Note: Individuals sensitive to latex should use vinyl or nitrile gloves.

3. REAGENT AND STANDARD PREPARATION

Note: ATP contamination will cause erroneous results and increase background. Wear gloves to prevent ATP contamination from your hands during reagent preparation and while performing the assay.

3.1 Equilibrate the sample buffer to room temperature.

3.2 Lightly tap the vial of the rL/L Reagent before opening to ensure the lyophilized material collects at the bottom of the vial.

3.3 Transfer the contents of the vial of rL/L Reconstitution Buffer to the vial of the rL/L Reagent.

3.4 Replace the stopper and slowly invert the vial several times to dissolve the contents. Do not shake the reagent bottle.

3.5 Allow the reconstituted rL/L Reagent to stand at room temperature for 1 hour before use.

Note: Reconstituted rL/L Reagent may be kept for 8 hours at room temperature. Store at -20°C in single-use aliquots for long-term storage.

3.6 Prepare a 10-fold serial dilution of the ATP standard (1x10-7 M) assay buffer in labeled test tubes. Dilute to 1x10-11 M ATP.

4. INSTRUMENT SET-UP

4.1 Turn on instrument and allow to warm up for at least 5 minutes (300 seconds).

4.2 Choose the Standard (STD) Mode and adjust settings so that:

Delay period = 2 sec.
Integrate period = 10 sec.
Replicates = 1

5. ATP STANDARD CURVE PROCEDURE

5.1 Prepare several test tubes with 100 µL of rL/L Reagent only.

5.2 Press <0> on the luminometer to begin the blank subtraction sequence.

5.3 Add 10 µL of assay buffer before pressing for each blank sample.

5.4 The luminometer will read and record the values for each blank sample. These values will be averaged and recorded as the background level.

5.5 Press <1> to accept the background level. This value will now be subtracted from subsequent measurements.

Note: It is important to determine the background level of ATP due to contamination. The blank subtraction should be performed each day the ATP assay is run.

5.6 From the dilutions obtained in step 3.6, add 10 µL of the lowest concentration (1x10-11 M) of ATP to a labeled test tube containing 100 µL rL/L Reagent.

5.7 Place the labeled test tube into the TD-20/20 Luminometer and close the lid.

5.8 Press to begin measurement. After the 2 second delay, the instrument will begin the 10-second measurement period.

5.9 Record this value and repeat with the next concentration of ATP. Continue for a total of five concentrations from 1x10-11 to 1x10-7 M ATP.

6. SAMPLE ANALYSIS

6.1 Add 10 µL of your sample prepared in the assay buffer to a labeled test tube containing 100 µL rL/L reagent.

6.2 Press . After the 2-second delay, the instrument will begin the 10-second measurement period.

6.3 Record this value and repeat steps 6.1 - 6.2 for your remaining samples.

6.4 Plot the RLU values of your samples along with the RLU values obtained during the standard curve procedure to determine the concentration of ATP in your samples.

7. ABOUT PROMEGA CORPORATION

Orders for Promega's products may be placed by:

Toll-Free: (800) 356-9526
Fax: (800) 356-1970
Email: custserv@promega.com

Mailing Address (HQ):
Promega Corporation
2800 Woods Hollow Rd.
Madison, WI 53711 USA


theLabRat.com 2005. All Rights Reserved.