The
Turner BioSystems TD-700 Laboratory Fluorometer in
combination with Molecular Probes' LIVE/DEAD®
BacLightTM Bacterial Viability Kit provides
a novel two-color fluorescence assay of bacterial
viability that allows researchers to quantitatively
distinguish live and dead bacteria in minutes, even
in a mixed population containing a range of bacterial
types. Conventional direct-count assays of bacterial
viability are based on metabolic characteristics or
membrane integrity. However, methods relying on metabolic
characteristics often only work for a limited subset
of bacterial groups,(1) and methods for assessing
bacterial membrane integrity commonly have high levels
of background fluorescence.(2) Both types of determinations
suffer from being very sensitive to growth and staining
conditions.(3,4)
The
LIVE/DEAD BacLight Bacterial Viability assay utilizes
mixtures of SYTO® 9 green fluorescent nucleic
acid stain and the red fluorescent nucleic acid stain,
propidium iodide. These stains differ both in their
spectral characteristics and in their ability to penetrate
healthy bacterial cells. When used alone, the SYTO
9 stain labels bacteria with both intact and damaged
membranes. In contrast, propidium iodide penetrates
only bacteria with damaged membranes, competing with
the SYTO 9 stain for nucleic acid binding sites when
both dyes are present. When mixed in recommended proportions,
SYTO 9 stain and propidium iodide produce green fluorescent
staining of bacteria with intact cell membranes and
red fluorescent staining of bacteria with damaged
membranes. The background remains virtually nonfluorescent.
Consequently, the ratio of green to red fluorescence
intensities provides a quantitative index of bacterial
viability (Figure 1).
Table
1. Preparation of isopropyl alcohol/water mixtures
3.1.
Disinfectant Treatment of Staphylococcus aureus with
Isopropyl Alcohol
3.1.1
In a set of 5 labeled tubes, prepare 0, 10, 30,
70 and 100% isopropyl alcohol/water mixtures as
outlined in Table 1.
3.1.2
Label a set of 5 microfuge tubes with the designation
0, 10, 30, 70, or 100%.
3.1.3
Transfer 1 mL of Staphylococcus aureus suspension
to each of the labeled microfuge tubes[A].
3.1.4
Centrifuge the microfuge tubes for 3 minutes at
10,000 rpm.
3.1.5
Remove and discard the supernatant from each microfuge
tube using a glass Pasteur pipet.
3.1.6
Break up the bacterial pellets and resuspend each
in 1 mL of 0, 10, 30, 70 or 100% isopropyl alcohol/water
(step 3.1.1) according to the designation on the
tube. For example, resuspend the pellet in the microfuge
tube marked 10% in 1 mL of 10% isopropyl alcohol.
3.1.7
Incubate the bacterial suspensions in alcohol/water
mixtures for 1 hour at room temperature.
3.1.8
Centrifuge the bacterial suspensions for 3 minutes
at 10,000 rpm.
3.1.9
Break up bacterial pellets and resuspend each in
1 mL of sterile filtered water.
3.1.10
Centrifuge bacterial suspensions for 3 minutes at
10,000 rpm.
3.1.11
Break up bacterial pellets and resuspend each in
1.5 mL of sterile filtered water.[A]
[A]
Although the experiment described here is based
on a comparison of samples containing equal numbers
of bacteria, it is also possible to compare viability
of bacterial samples with different densities because
the ratio of green to red fluorescence is relatively
independent of the number of bacteria present.
3.2
Bacterial Staining
3.2.1
Prepare a 2X BacLight staining solution from the
kit reagents by mixing 3 uL of component A and 3
uL of component B per 1 mL of sterile water. To
prepare sufficient staining solution for the five
samples and one reagent blank used in the experiment
described here, add 27 uL of BacLight component
A and 27 uL of BacLight component B to 9 mL of sterile
water.
3.2.2
Label six 13 mm glass test tubes with reagent blank,
0, 10, 30, 70, and 100%.
3.2.3
Transfer 1.5 mL of 2X BacLight (prepared in 3.2.1)
to each labeled test tube.
3.2.4
Transfer 1.5 mL of sterile water to the reagent
blank tube, making up a total of 3 mL in this test
tube.
3.2.5
Transfer 1.5 mL of the isopropyl alcohol-exposed
Staphylococcus aureus from the microfuge tubes to
the correspondingly labeled glass test tubes, making
up a total of 3 mL in each test tube.
3.3
Fluorescence Measurements (Green Fluorescence)
3.3.1
Install the 486 nm filter in A-EX and 520 nm filter
in A-EM.
3.3.2
Rotate the TD-700 filter cylinder to place the 485
nm excitation and 530 nm emission filters in the
optical path.
3.3.3
You will be performing a simple mode calibration
(refer to your manual if need be). Press "enter"
from the keypad. Select [1] on the keypad to enter
"setup" then select [1] to enter "mode".
3.3.4
Using the arrow key to choose the mode, select "simple,
then [ESC] twice to return to the setup/calibration
screen. Select [2], then press [ENTER].
3.3.5
The TD-700 will prompt you to insert a typical sample.
Insert the sample containing the largest number
of live bacteria; this sample should have the highest
intensity of green fluorescence among those to be
measured. In the experiment described here, this
will be the sample containing 0% isopropyl alcohol.
3.3.6
Press [ENTER]. The instrument detection sensitivity
will be automatically optimized to match the fluorescence
of the sample. When the screen indicates that the
sample is 500, press [ENTER]. Your TD-700 is now
calibrated.
3.3.7
Press [ENTER] again and read the remaining samples
in the order of reagent blank followed by the bacterial
samples in order of increasing isopropyl alcohol
concentration.
3.3.8
For each sample, record the fluorescence intensity
value shown on the fluorometer display screen. To
equalize any photobleaching effects, insert samples
into the fluorometer for approximately equal time
periods. Correct data for the dye background
by subtracting the reagent blank fluorescence value
from each bacterial sample fluorescence reading.
3.4
Fluorescence Measurements (Red Fluorescence)
3.4.1
Rotate the TD-700 filter cylinder to place the 520
nm excitation and 325-700 nm and greater than 610
nm emission filters in the optical path ("B"
on the left hand side lined up with silver dot).
3.4.2
Press [ENTER] on the TD-700 key pad.
3.4.3
Select [2] CALIBRATION from the displayed options.
3.4.4
Insert the sample containing the largest number
of dead bacteria; this sample should have the highest
intensity of red fluorescence among those to be
measured. In the experiment described here, this
will be the sample containing 100% isopropyl alcohol.
3.4.5
Press ENTER. The instrument detection sensitivity
will be automatically optimized to match the fluorescence
of the sample.
3.4.6
Press ENTER again and read the remaining samples
in the order of reagent blank followed by the bacterial
samples in order of increasing isopropyl alcohol
concentration.
3.4.7
For each sample, record the fluorescence intensity
value shown on the fluorometer display screen. To
equalize any photobleaching effects, insert samples
into the fluorometer for approximately equal time
periods. Correct data for the dye background
by subtracting the reagent blank fluorescence value
from each bacterial sample fluorescence reading.
3.5
Data Analysis and Interpretation
3.5.1
For each sample, divide the background-corrected
green fluorescence intensity by the background-corrected
red fluorescence intensity.
3.5.2
Plot the calculated green/red fluorescence intensity
ratio (Y-axis) as a function of isopropyl alcohol
concentration (X-axis).
3.5.3
Plotted data should show decreasing green/red fluorescence
intensity ratios as the number of dead bacteria
increases (i.e. as the isopropyl alcohol concentration
increases; Figure 1).